Dew-point instrument TMM-100

This equipment is the dew-point instrument which measures the dew-point temperature and dry-bulb temperature, and consists of the dew-point detection unit and the dew-point temperature measurement unit.
The dew-point temperature measurement unit can digitally display the dew-point temperature and dry-bulb temperature detected by the sensor and output the value as an analog voltage signal.
Since the detection unit and the dew-point measurement unit can be installed separately, it is suitable for applications such as checking the condition of another room where the detection unit is installed.

TMM-100 Specs

Dew-point sensor

Thermal average temperature measurement method by saturated vapor pressure of lithium chloride

Operating temperature range -45 to +60℃
Detection Gold-plating wire heater
Resistance temperature sensor Nickel wire
Wind velocity Less than 0.25m/sec
Heater power supply 25V AC

Dry-bulb temperature sensor

Detection element Resistance temperature sensor JIS Pt100
Operating temperature range -50 to +80℃

Dew-point temperature measurement unit

Power supply AC100V±10% 50/60Hz Approx.150VA
Environmental conditions
Temperature
  0to 40℃
Humidity
20to 80%RH (No dew condensation)
Display element Comprehensive accuracy
(The accuracy of the sensor is not included.)
Dew-point temperature
Accuracy rating ±0.4℃
Temperature drift ±0.2℃/10℃
Dry-bulb temperature
Accuracy rating ±0.3℃
Temperature drift ±0.2℃/10℃
Analog output Number of outputs There are two outputs each for Dew-point temperature and Dry-bulb
temperature, totaling 4 outputs

Output voltage
Dew-point temperature
-4.5 to +6.0V / -45.0 to +60.0℃
Dry-bulb temperature
-5.0 to +8.0V / -50.0 to +80.0℃

Comprehensive accuracy(The accuracy of the sensor is not included.)
Dew-point temperature
Accuracy rating ±0.6% of F.S.
Temperature drift ±0.5%/10℃
Dry-bulb temperature
Accuracy rating ±0.35% of F.S.
Temperature drift ±0.5%/10℃

Test room and Measurement room cable

Electric wire 1.25sq Vinyl cab tire electric wire 3 set
Cable length Production according to installation environment

Option

Desktop case

A case is used to enable the installation at a location where 19-inch racks are not available, or at a desired location.
Dimensions:500(W)×220(H)×453(D)mm (Not including protrusions)
※ It is possible to make the case fit the customer’s usage environment.
(With castors etc.)

Signal output cable

Is the cable for TMM-100 signal outputting to a peripheral device.
We can manufacture according to the customer’s system.

Application example

Application example of TMM-100

  • Environmental evaluation of automobile
  • Automobile running mode exhaust gas test
  • Running mode fuel consumption rate test of automobile
  • Various environmental tests